Configure Display Variable 1 to track the primary mA Output
Display
ProLink III
Field Communicator Not available
Overview
You can configure Display Variable 1 to track mA Output Process Variable for the primary
mA Output. When tracking is enabled, you can control Display Variable 1 from the display
menu.
Tip
This feature is the only way to configure a display variable from the display menus, and it applies only
to Display Variable 1.
Procedure
Configure Display Variable 1 to track the primary mA Output.
Display Variable 1 will automatically be set to match mA Output Process Variable for the
primary mA Output. If you change the configuration of mA Output Process Variable,
Display Variable 1 will be updated automatically.
5.1.3
Configure the number of decimal places (precision)
shown on the display
Display
ProLink III
Field Communicator Configure > Manual Setup > Display > Decimal Places
Overview
You can specify the number of decimal places (precision) that are shown on the display for
each process variable or diagnostic variable. You can set the precision independently for
each variable.
The display precision does not affect the actual value of the variable or the value used in
calculations.
Procedure
1.
2.
Configuration and Use Manual
OFF-LINE MAINT > OFF-LINE CONFG > DISPLY > VAR 1
Device Tools > Configuration > Transmitter Display > Display Security
Not available
Device Tools > Configuration > Transmitter Display > Display Variables
Select a variable.
Set Number of Decimal Places to the number of decimal places you want shown
when the process variable or diagnostic variable appears on the display.
Configure device options and preferences
69