Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density
Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value
configured for Standard Gas Density or polled value if configured for polled base density.
4.6
Configure temperature measurement
The temperature measurement parameters control how temperature data from the
sensor is reported.
4.6.1
Configure Temperature Measurement Unit
Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature
measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit.
Different communications tools may use different labels for the units.
Unit description
Degrees Celsius
Degrees Fahrenheit
Degrees Rankine
Kelvin
Configuration and Use Manual
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP
Device Tools > Configuration > Process Measurement > Temperature
Display
°C
°F
°R
°K
Configure process measurement
Label
Field Communica-
ProLink III
tor
°C
degC
°F
degF
°R
degR
°K
Kelvin
47