Configure Pressure Compensation Using The Field Communicator - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Configure process measurement
10.
Postrequisites
If you are using an external pressure value, verify the setup by checking the External
Pressure value displayed in the Inputs area of the main window.
4.9.2
Configure pressure compensation using the Field
Communicator
1.
2.
3.
4.
5.
6.
64
If you want to use digital communications, click Apply, then perform the necessary
host programming and communications setup to write pressure data to the
transmitter at appropriate intervals.
Choose Online > Configure > Manual Setup > Measurements > External Pressure/
Temperature > Pressure.
Set Pressure Compensation to Enabled.
Enter Flow Cal Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and
defines the pressure at which there is no pressure effect. If the data is unavailable,
enter 20 PSI.
Enter Flow Press Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the
value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
Enter Dens Press Factor for your sensor.
The density factor is the change in fluid density, in g/cm
value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm
Determine how the transmitter will obtain pressure data, and perform the required
setup.
Option
Setup
A user-configured
a. Set Pressure Unit to the desired unit.
static pressure value
b. Set Compensation Pressure to the desired value.
Micro Motion Model 2700 Transmitters with Intrinsically Safe Outputs
3
/PSI. When entering the
3
/PSI, enter −0.000006g/cm3/PSI.