3.2

Default values and ranges

See
Section D.1
parameters.
3.3
Enable access to the off-line menu of the
display
Display
ProLink III
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
3.4
Disable write-protection on the transmitter
configuration
Display
ProLink III
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it
before you can change any configuration parameters. By default, the transmitter is not
write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent
normal operational use. You can always disable write-protection, perform any required configuration
changes, then re-enable write-protection.
Configuration and Use Manual
to view the default values and ranges for the most commonly used
OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
Device Tools > Configuration > Transmitter Display > Display Security
OFF-LINE MAINT > CONFG > LOCK
Device Tools > Configuration > Write-Protection
Introduction to configuration and commissioning
15