For T-Series sensors only, you have the option to perform a D3 and D4 calibration to improve the
accuracy of the density measurement if the fluid density is outside of the 0.8 g/cm
range. If you choose to perform the D3 and D4 calibration, do not perform the D1 and D2 calibration.
Procedure
To perform a D1 and D2 calibration, see Figure 9-2 .
Figure 9-2
D1 and D2 density calibration using Field Communicator
9.4
Perform a D3 and D4 density calibration (T-Series
sensors only)
For T-Series sensors, the optional D3 and D4 calibration could improve the accuracy of the density
measurement if the fluid density is outside of the 0.8 g/cm
Configuration and Use Manual
3
to 1.2 g/cm
3
density range.
Measurement support
3
to 1.2 g/cm
3
density
171