Configure The Frequency Output; Configure Frequency Output Process Variable - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Options for mA Output Fault Action and mA Output Fault
Level
Option
Upscale
Downscale (default)
Internal Zero
None
6.3

Configure the Frequency Output

The Frequency Output is used to report a process variable. The Frequency Output
parameters control how the process variable is reported. Your transmitter may have zero
or one Frequency Output: Channel B can be configured as a Frequency Output or a
Discrete Output. The default assignment for Channel B is Frequency Output.
Important
Whenever you change a Frequency Output parameter, verify all other Frequency Output parameters
before returning the flowmeter to service. In some situations, the transmitter automatically loads a
set of stored values, and these values may not be appropriate for your application.
6.3.1

Configure Frequency Output Process Variable

Display
ProLink III
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > Third
Overview
Frequency Output Process Variable controls the variable that is reported over the
frequency output.
Configuration and Use Manual
mA Output behavior
Goes to the configured fault level
Goes to the configured fault level
Goes to the mA Output level associated
with a process variable value of 0 (zero),
as determined by Lower Range Value and
Upper Range Value settings
Tracks data for the assigned process vari-
able; no fault action
OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO SRC
Device Tools > Configuration > I/O > Outputs > Frequency Output
Variable
Integrate the meter with the control system
mA Output Fault Level
Default: 22.0 mA
Range: 21.0 to 24.0 mA
Default: 3.2 mA
Range: 3.2 to 3.6 mA
Not applicable
Not applicable
97