1-34. AC Current Specifications

Range
Resolution
220 µA
2.2 mA
22 mA
220 mA
2.2 A
5725A Amplifier:
11 A
Note:
Maximum output from the calibrator's terminals is 2.2 A. Uncertainty specifications for 220 µA and 2.2 mA ranges are increased by a factor
of 1.3 plus 2 µA when supplied through 5725A terminals. For the 5720A 220 µA range, 1 kHz through 5 kHz and 5 kHz through 10 kHz,
when the output is coming from the AUX current terminal, use the 5700A Absolute Uncertainty Specifications.
otherwise identical for all output locations.
For fields strengths >0.4 V/m but ≤3 V/m, add 1 % of range.
1.
1-36
5720A Series II AC Current Specifications: 99 % Confidence Level
Frequency
24 Hours
(Hz)
10 - 20
260 + 20
20 - 40
170 + 12
1 nA
40 - 1 k
120 + 10
1k - 5 k
300 + 15
5k - 10 k
1000 + 80
10 - 20
260 + 50
20 - 40
170 + 40
10 nA
40 - 1 k
120 + 40
1k - 5 k
210 + 130
5k - 10 k
1000 + 800
10 - 20
260 + 500
20 - 40
170 + 400
100 nA
40 - 1 k
120 + 400
1k - 5 k
210 + 700
5k - 10 k
1000 + 6000
10 - 20
20 - 40
1 µA
40 - 1 k
1k - 5 k
5k - 10 k
1000 + 12
20 - 1 k
290 + 40
10 µA
1 k - 5 k
440 + 100
5 k - 10 k
6000 + 200
40 - 1 k
370 + 170
100 µA
1 k - 5 k
800 + 380
5 k - 10 k
3000 + 750
Absolute Uncertainty
±5 °C from calibration temperature
90 Days
180 Days
± (ppm output + nA)
280 + 20
290 + 20
180 + 12
190 + 12
130 + 10
135 + 10
320 + 15
340 + 15
1100 + 80
1200 + 80
280 + 50
290 + 50
180 + 40
190 + 40
130 + 40
135 + 40
220 + 130
230 + 130
1100 + 800
1200 + 800
280 + 500
290 + 500
180 + 400
190 + 400
130 + 400
135 + 400
220 + 700
230 + 700
1100 + 6000
1200 + 6000
± (ppm output + µA)
260 + 5
280 + 5
290 + 5
170 + 4
180 + 4
190 + 4
120 + 3
130 + 3
135 + 3
210 + 4
220 + 4
230 + 4
1100 + 12
1200 + 12
300 + 40
310 + 40
460 + 100
480 + 100
7000 + 200
7500 + 200
400 + 170
440 + 170
850 + 380
900 + 380
3300 + 750
3500 + 750
Relative Uncertainty
[1]
±1 °C
1 Year
24 Hours
300 + 20
260 + 20
200 + 12
130 + 12
140 + 10
100 + 10
350 + 15
250 + 15
1300 + 80
900 + 80
300 + 50
260 + 50
200 + 40
130 + 40
140 + 40
100 + 40
240 + 130
250 + 130
1300 + 800
900 + 800
1000 + 800
300 + 500
260 + 500
200 + 400
130 + 400
140 + 400
100 + 400
240 + 700
250 + 700
1300 + 6000
900 + 6000
1000 + 6000
300 + 5
260 + 5
200 + 4
130 + 4
140 + 3
100 + 3
240 + 4
250 + 4
1300 + 12
900 + 12
320 + 40
300 + 40
500 + 100
500 + 100
8000 + 200
6000 + 200
7000 + 200
460 + 170
300 + 170
950 + 380
700 + 380
3600 + 750
2800 + 750
3200 + 750
Specifications are
90 Days
280 + 20
150 + 12
110 + 10
280 + 15
1000 + 80
280 + 50
150 + 40
110 + 40
280 + 130
280 + 500
150 + 400
110 + 400
280 + 700
280 + 5
150 + 4
110 + 3
280 + 4
1000 + 12
350 + 40
520 + 100
330 + 170
800 + 380