Ac Voltage, Square Wave Characteristics; Ac Voltage, Triangle Wave Characteristics (Typical); Ac Current (Non-Sine Wave) - Fluke 5502E Getting Started

Multi-product calibrator
Table of Contents

Advertisement

5502E
Getting Started

AC Voltage, Square Wave Characteristics

Risetime @
Settling Time @
1 kHz
Typical
<1 μs

AC Voltage, Triangle Wave Characteristics (typical)

0.3 % of p-p value, from 10 % to 90 % point

AC Current (Non-Sine Wave)

Triangle Wave &
Truncated Sine Wave
Range p-p
0.047 to 0.92999 mA
0.93 to 9.29999 mA
9.3 to 92.9999 mA
93 to 929.999 mA
0.93 to 8.49999 A
8.5 to 57 A
[1]
Frequency limited to 1 kHz with LCOMP on.
[2]
Frequency limited to 440 Hz with LCOMP on.
24
Overshoot
@ 1 kHz
1 kHz Typical
Typical
<10 μs to 1 % of
<2 %
final value
Linearity to 1 kHz
Frequency
10 to 45 Hz
[1]
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
[1]
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
[1]
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
[1]
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
[2]
45 Hz to 1 kHz
1 to 10 kHz
45 to 500 Hz
[2]
500 Hz to 1 kHz
Duty Cycle Range
±(0.02 % of period + 100 ns), 50 % duty cycle
1 % to 99 % <3.3 V p-p.
±(0.05 % of period + 100 ns), other duty cycles
0,01 Hz to 100 kHz
<1 % of p-p value, with amplitude >50 % of range
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.5
10 + 2
0.5 + 1.0
0.5 + 0.5
10 + 2
0.5 + 0.5
1.0 + 1.0
Duty Cycle Uncertainty
from 10 % to 90 %
Aberrations
Max Current
Resolution
Six digits
Six digits
Six digits
Six digits
Six digits

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents