Exchanging The Nano Pump Main Board (Npm Board) - Agilent Technologies 1100 Series Service Manual

Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Repairing the Pump

Exchanging the Nano Pump Main Board (NPM Board)

Frequency
Tools required
Parts required
1 Turn off the pump, disconnect all cables and remove the pump from the
2 Remove the top covers and foam, see "Removing the Top Cover and Foam"
3 Disconnect all connectors from the main board.
Connector
S1
J1
J2
J3
J4
J5
J6
P2
P8
P12
P14
When removing connectors, counterhold on connector J3 with one hand.
NO TE
134
Board defective
Wrench 14 mm
Wrench 7 mm
Wrench 5 mm
NPM board, G2226-66530, exchange part number G2226-69530
stack.
on page 131.
Description
Configuration switch
GPIB
CAN connector
CAN connector
RS232
Remote
Analog output
Power supply
Motor B
Motor A
Encoder A
Connector
Description
P11
AIV - A
P15
Leak sensor
P16
Damper
P17
Fan
P18
Interface board
P21
Encoder B
P20
AIV-B
P22
Solvent Selection Valve
P9
EMPV
P19
Flow sensor
1100 Series Nano Pump Service Manual