Control Module Screens for the Nano Pump
6
Firmware Update
Use the Esc-key to receive Views on the F5-key. Choose System from the
pull-down menu. Use the F3-key (Records) to select the pump. Use the F5-key
(FW Update) to enter the Update section. If you want to update the resident
firmware (together with specific main firmware revisions), select the file from
the PCMCIA card (RESnnnn.DLB) and press execute. If you want to update
the main firmware, press F7-key (Transfer) to turn the module into the
resident mode (LED on module should blink yellow).
Use the Esc-key to receive Views on the F5-key. Choose System from the
pull-down menu. Use the F3-key (Records) to select the Generic module. In
this screen the resident firmware revision is shown.
Use the F5-key (FW Update) to enter the Update section. Select the a file from
the PCMCIA card (1376nnnn.DLB) and press execute. When the update has
finished, press F7-key (Transfer) to return the module into the normal mode
(LED on module should stay yellow).