Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Troubleshooting and Test Functions
The slope and plateau is evaluated automatically. "Evaluating the Results" on
page 71 describes the evaluation and interpretation of the pressure test
results
Running the test from the Control Module
1 Place a bottle of LC-grade isopropyl alcohol in the solvent cabinet and
2 Block the injection device (port 6) with a blank nut (01080-83202),
3 Set flow for channel A2 to 2500 µl/min and flush the degasser channel for
4 Set flow to 0 µl/min.
5 Connect the signal cable to the analog output at the rear of the nano pump
6 Press Execute to initialize the pressure test.
70
connect it to channel A2.
"Positioning the blank nut"
about 5 minutes.
(only if an integrator is used).
on page 69
1100 Series Nano Pump Service Manual