Agilent Technologies 1100 Series Service Manual page 136

Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Repairing the Pump
Seal Wear-in Procedure
This procedure is required for standard seals only (5063-6589), but it will definitely damage
CAU TI O N
the normal phase application seals (0905-1420).
1 Fill a bottle with 100 ml Isopropanol and place the bottle head assembly of
2 Screw the adapter (0100-1847) to the AIV and connect the inlet tube from
3 Connect the restriction capillary (5022-2159) to the outlet of the EMPV.
4 Turn the system in purge mode and purge the system for 2 minutes with
5 Turn the system to standard mode, set the flow to a rate adequate to achieve
6 Turn OFF the pump, slowly disconnect the restriction capillary from the
7 Rinse your system with the solvent used for your next application.
124
the pump head that is supposed to be worn-in into the bottle.
the bottle head directly to it.
Insert its other end into a waste container.
isopropanol at a flow rate of 2 ml/min.
a pressure of 350 bar. Pump 15 minutes at this pressure to wear in the
seals. The pressure can be monitored at your analog output signal, with the
handheld controller, Chemstation or any other controlling device connected
to your pump.
EMPV to release the pressure from the system. Reconnect the capillary
going to the flow sensor and the connecting tube from solvent selection
valve to the AIV.
1100 Series Nano Pump Service Manual