Repairing the Pump
8 Place the new fan into the recess. Make sure that the air flow is directed as
9 Re-install pump assembly B. Reconnect all capillaries and connectors. (P8,
10 Re-install the Z-panel and the damper. Reconnect the capillaries and
11 Re-install the mixing chamber to its holder.
12 Re-install the flow sensor. Reconnect the capillaries and the cable.
13 Re-install the EMPV. Reconnect the capillaries, the waste tube and the
14 Replace the top foam section, optional interface board, front cover, and top
144
indicated (arrow on the fan points in the same direction as on the
foam).Connect the cable to the main board (P17).
P21).
connectors. (P16, P11, P20)
cable.
cover (see "Replacing the Top Cover and Foam" on page 154).
1100 Series Nano Pump Service Manual