Troubleshooting and Test Functions
.
Micro Mode Pressure Test Results
The test results are evaluated automatically. The sum of all leaks within the
column flow path from the EMPV to the blank nut must be lower
than100 nl/min.
Small leaks, with no visible leaks in the flow path can cause the test to fail.
NO TE
66
1100 Series Nano Pump Service Manual