Removing And Disassembling The Pump Head Assembly - Agilent Technologies 1100 Series Service Manual

Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Repairing the Pump

Removing and Disassembling the Pump Head Assembly

Never start the pump when the pump head is removed. This may damage the pump
CAU TI O N
drive.
When required:
Exchanging the seals
Exchanging the plungers
Exchanging seals of the seal wash option
Tools required:
Wrench 1/4 inch
3-mm hexagonal key
4-mm hexagonal key
Preparations for this procedure:
Switch off nano pump at power switch
Remove the front cover to have access to
the pump mechanics
120
1 Disconnect the capillary at the pumphead adapter
and the tube at the active inlet valve. Beware of
leaking solvents. Disconnect the active inlet valve
cable plug.
Active inlet
valve
Plug
1100 Series Nano Pump Service Manual