6
Control Module Screens for the Nano Pump
Screens available from the Records screen
Records screen
Use the Esc-key until you receive Views on the F5-key. Choose System from
the pull-down menu. Use the F4-key (Records) and select the pump. Errors
are reported either into the F2 (System Log) or F3 (Error Log).
EMF
Use the F1-key (EMF) to set EMF Limits. Choose menu item Setup Limits to
Early Mainte-
set the amount of pumped solvent or seal wear, when you want to receive a
nance Feedback)
warning for exceeded limits.
Press F7 (Reset) to reset the counters after exchanging parts that are subject
to wear. Changes must be activated with F6 (Done).
248
1100 Series Nano Pump Service Manual