Repairing the Pump
4 Slide the pump head assembly onto the pump drive.
Apply a small amount of pump head grease to the
pumphead screws and the balls of the spindle drive.
Tighten screws stepwise with increasing torque.
128
5 Reconnect the capillaries, tubing and the active inlet
Balls of
spindle drive
Pump head
screws
valve cable to the connector.
Active inlet valve