Agilent Technologies MSO6102A Programmer's Reference Manual page 96

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem
Table 37 Status Byte Register (STB) (continued)
Bit
6
5
4
3
2
1
0
To read the instrument's status byte with RQS reported on bit 6, use the interface Serial
N O T E
Poll.
See Also
96
Name
Description
RQS
Request Service
MSS
Master Summary Status
ESB
Event Status Bit
MAV
Message Available
---
---
MSG
Message
USR
User Event
TRG
Trigger
"Introduction to Common (*) Commands"
"*SRE (Service Request Enable)"
When Set (1 = High = True), Indicates:
When polled, that the device is requesting service.
When read (by *STB?), whether the device has a
reason for requesting service.
An enabled condition in the Standard Event Status
Register (ESR) has occurred.
There are messages in the Output Queue.
(Not used, always 0.)
An advisory has been displayed on the
oscilloscope.
An enabled user event condition has occurred.
A trigger has occurred.
on page 77
on page 93
Agilent 6000 Series Oscilloscopes Programmer's Reference

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents