Agilent Technologies MSO6102A Programmer's Reference Manual page 23

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

Command
:TRIGger:LIN:SAMPlepoint (see
:TRIGger:LIN:STANdard (see
:TRIGger:LIN:SYNCbreak (see
:HWEenable (see
:HWERegister:CONDition? (see
:HWERegister[:EVENt]? (see
Changed
Commands
Command
:SBUS:MODE (see
:TRIGger:CAN:TRIGger (see
:TRIGger:LIN:TRIGger (see
Obsolete
Commands
Obsolete Command
:TRIGger:CAN:SIGNal:DEFiniti
on (see
:TRIGger:LIN:SIGNal:DEFinitio
n (see
Agilent 6000 Series Oscilloscopes Programmer's Reference
page
page
page
page
113)
page
page
page
310)
page
352)
page
407)
Current Command Equivalent
none
page
529)
none
page
530)
Description
402)
Sets the point during the bit time where the bit
level is sampled to determine whether the bit is
dominant or recessive. The sample point
represents the percentage of time between the
beginning of the bit time to the end of the bit
time.
405)
Sets the LIN standard in effect for triggering
and decoding to be LIN1.3 or LIN2.0.
406)
Sets the length of the LIN sync break to be
greater than or equal to 11,12, or 13 clock
lengths. The sync break is the idle period in the
bus activity at the beginning of each packet
that distinguishes one information packet from
the previous one.
Sets or reads the hardware event enable mask
register.
115)
Queries the hardware event condition register.
117)
Queries the hardware event event register.
Differences
The CAN and LIN serial bus decode modes
have been added.
The ALLerrors, OVERload, and ACKerror
options have been added.
The ID (for Frame Id) option has been added.
What's New
Behavior Differences
1
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents