The :ACQuire:TYPE AVERage command sets the oscilloscope in the
averaging mode. You can set the count by sending the :ACQuire:COUNt
command followed by the number of averages. In this mode, the value for
averages is an integer from 1 (smoothing) to 65536. The COUNt value
determines the number of averages that must be acquired.
Peak Detect
The :ACQuire:TYPE PEAK command sets the oscilloscope in the peak
detect mode. In this mode, :ACQuire:COUNt has no meaning.
Real-time Mode
The :ACQuire:MODE RTIMe command sets the oscilloscope in real- time
mode. This mode is useful to inhibit equivalent time sampling at fast
sweep speeds.
Equivalent-time Mode
The :ACQuire:MODE ETIME command sets the oscilloscope in
equivalent- time mode.
Reporting the Setup
Use :ACQuire? to query setup information for the ACQuire subsystem.
Return Format
The following is a sample response from the :ACQuire? query. In this case,
the query was issued following a *RST command.
:ACQ:MODE RTIM;TYPE NORM;COMP 100;COUNT 8