Table 38 Root (:) Commands Summary (continued)
Command
:DIGitize
[<source>[,..,<source>]
] (see
page
111)
:HWEenable <n> (see
page
113)
n/a
n/a
:MERGe <pixel memory>
(see
page
119)
:OPEE <n> (see
page
120)
n/a
n/a
:OVLenable <mask> (see
page
126)
n/a
Agilent 6000 Series Oscilloscopes Programmer's Reference
Query
n/a
:HWEenable? (see
page
113)
:HWERregister:CONDition
? (see
page
115)
:HWERegister[:EVENt]?
(see
page
117)
n/a
:OPEE? (see
page
121)
:OPERregister:CONDition
? (see
page
122)
:OPERegister[:EVENt]?
(see
page
124)
:OVLenable? (see
page
127)
:OVLRegister? (see
page
128)
Commands by Subsystem
Options and Query Returns
<source> ::= {CHANnel<n> | FUNCtion
| MATH | SBUS} for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | POD{1 | 2} |
BUS{1 | 2} | FUNCtion | MATH | SBUS}
for MSO models
<source> can be repeated up to 5
times
<n> ::= 1-2 or 1-4 in NR1 format
<n> ::= 16-bit integer in NR1 format
<n> ::= 16-bit integer in NR1 format
<n> ::= 16-bit integer in NR1 format
<pixel memory> ::= {PMEMory{0 | 1 |
2 | 3 | 4 | 5 | 6 | 7 | 8 | 9}}
<n> ::= 16-bit integer in NR1 format
<n> ::= 16-bit integer in NR1 format
<n> ::= 16-bit integer in NR1 format
<mask> ::= 16-bit integer in NR1
format as shown:
Bit Weight Input
--- ------ ----------
10
1024
Ext Trigger Fault
9
512
Channel 4 Fault
8
256
Channel 3 Fault
7
128
Channel 2 Fault
6
64
Channel 1 Fault
4
16
Ext Trigger OVL
3
8
Channel 4 OVL
2
4
Channel 3 OVL
1
2
Channel 2 OVL
0
1
Channel 1 OVL
<value> ::= integer in NR1 format.
See OVLenable for <value>
3