● Load cell manufacturer
service case.
● Load cell order number
service case.
7.3.5

Complete automatic calibration

● The scale must be empty (only mechanical dead load).
● Activate the "Automatic Calibration (82)" command.
7.3.6

Receive all data

Activate the "Receive all data" function in the communication menu.
All parameters can now been saved as a backup file on the hard disk. If a module is
replaced, the backup file can be downloaded to the new module within a few seconds. At the
time of input of the backup file, the scale is directly in the calibrated state again – without a
new calibration.
7.3.7

Checking the scale after calibration

Perform the following steps:
1. Scale is unloaded and shows "0 kg".
2. Place a known reference weight on the scale.
→ Check the displayed value.
3. If a second known reference weight is available, place it on the scale additionally.
→ Check whether the scale displays the sum of the reference weights.
4. Remove the reference weight from the scale.
→ Check that the display is "0 kg" again.
Should these steps not yield the expected results, the electrical connection of the load cells
and the scale mechanics must be checked.
SIWAREX WP251
Manual, 12/2015, A5E37203357A
The manufacturer of the installed load cells can be quickly and easily identified in a
The order number of the installed load cells can be quickly and easily identified in a
7.3 Automatic quick calibration with SIWATOOL
Commissioning
45

Advertisement

Table of Contents
loading

Table of Contents