Errors and messages
9.6 Operating errors
Technology errors
2003 Initial zero setting on not
possible
2004 Trace memory full
Reserve
Reserve
Reserve
2101 Risk of overfilling
2102 Coarse shut-off point al-
ready exceeded
Reserve
2104 CPU in "STOP"
2105 Set point too small
2106 Error fine weight
2107 Stop after tolerance error
Reserve
2109 Blocking time "coarse"
error
2110 Blocking time "coarse"
error
2111 Maximum weighing time
exceeded
2112 Logging not possible - no
standstill
2113 Maximum emptying time
exceeded
166
Error code Description and remedy
2003
The weight at switch-on is outside the configured
value range in data record DR 3 defined by the maxi-
mum positive and negative weight for initial zero set-
ting.
2004
The trace recording has been aborted because the
internal trace memory is full.
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
2101
Current parameter assignment would lead to overfill-
ing of the scale during dosing. Check the specified set
point in DR 20 or the limits for taring / zero setting in
DR 3 and DR 25.
2102
The current weight is already above the coarse shut-
off point. Remedy: Adjust set point or empty scale.
Reserve
Reserve
2104
WP251 is not in stand-alone operation, the SIMATIC
CPU is in STOP state and a dosing command has
been issued. Remedy: Put the SIMATIC CPU in RUN
state or activate stand-alone operation of WP251.
NOTICE: Ensure the safety of the system before acti-
vating stand-alone operation because WP251 acti-
vates dosing elements if necessary despite the
stopped SIMATIC CPU.
2105
The currently specified set point is too small.
Remedy: Check the set point setting in DR 20.
2106
Trailing weight (DR 23) and set point (DR 20) are
incompatible. Check the two entries.
2107
The weighing cycle was stopped due to a tolerance
error according to the parameter assignment in DR 22.
Reserve
Reserve
2109
After expiration of the blocking time coarse signal (DR
23), the coarse shut-off point has already been ex-
ceeded. Remedy: Decrease the blocking time coarse
signal in DR 23.
2110
After expiration of the blocking time fine signal (DR
23), the fine shut-off point has already been exceeded.
Remedy: Decrease the blocking time fine signal in DR
23.
2111
The maximum weighing time specified in DR 25 was
exceeded.
2112
The conditions required for automatic logging (e.g.
standstill 2) were not present at the time of the log
print. Remedy: Check the standstill 2 criteria in DR 3
and adapt it as necessary.
2113
The maximum emptying time configured in DR 25 was
exceeded during emptying. Remedy: Check the scale
mechanics (for possible soiling). Adjust the maximum
emptying time in DR 25, if necessary.
SIWAREX WP251