Errors and messages
9.6 Operating errors
Data and command errors
6053 Calibration command
error
6054 Scale command error
6055 Weighing command
error
168
Error code
Additional
string
4519
4522
6053
4510
4511
4520
4521
4522
4523
4524
4527
6054
4500
4501
4502
4530
4531
6055
4500
4501
4502
4508
4509
4513
4515
4540
4542
4543
4544
4545
4546
4547
4548
4549
Description
, because no test operation (only relevant in test field)
, because calibration not yet complete
Command from the Calibration commands group could not be
executed. Additional information contains details.
, because service mode is not active
, because write protection active
, because increment between calibration digits is too small
, because order of calibration points is incorrect
, because calibration not yet complete
, because calibration digits not in the permissible range
, because calibration weight 0
, because command is not permitted in simulation mode
Command from the Scale commands group (Zero, Tare, etc.)
could not be executed. Additional information contains details.
, because a fault is present
, because no standstill occurred
, because module is already waiting for standstill
, because weight is outside the permissible tare weight range
, because weight is outside the permissible zero setting range
Command from the Weighing commands group (Start single
weighing, Start continuous mode, etc.) could not be executed.
Additional information contains details.
, because a fault is present
, because no standstill occurred
, because module is already waiting for standstill
, because dosing is not active
, because timing within dosing step sequence is impermissible
, because service mode is activated
, because not possible in this operating mode
, because weight above fine shut-off point
, because set point too small
, because set point too large
, because tolerance limits implausible
, because fine weight, trailing weight and set point are incompati-
ble
, because dosing would cause overfilling
, because configured emptying implausible
, because no check stop points defined
, because scale is already empty