7.5.2

Windows and functions of SIWATOOL

Control elements for SIWATOOL and operation of the scale
Parameter list of the SIWATOOL module
Error message buffer with time stamp

Image 7-5

Layout of the SIWATOOL user interface
Sending/receiving a data record is always carried out by right-clicking on the data record
name in the "Value" column in the tree structure.
For example, if data record 3 is to be sent, right-click on "Calibration parameter (DR3)". A
submenu is then opened with the option to send the respective data record to the weighing
module or read it from the module. All data records can only be sent to or read from the
SIWAREX as complete packets. It is not possible to read or write individual parameters
within a data record. Therefore the complete data record must initially be received for every
change to parameters within it. The desired parameter can then be edited, and the data
record returned. If the data record is not received, the danger exists that different offline
parameters will be sent to the scale and overwrite previously active and intentionally defined
parameters.
SIWAREX WP251
Manual, 12/2015, A5E37203357A
7.5 Service with the SIWATOOL program
Offline values of the SIWAREX module
Online values of the connected SIWAREX
module
Commissioning
49