9190A Input Panel Calibration; Fundamentals; Terminology - Fluke 9190A Operator's Manual

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A Input Panel Calibration

For assistance with the calibration process, contact an Authorized Service
Center. See "How to Contact Fluke Calibration" in Chapter 1.
This procedure is to be considered a general guideline. Each laboratory
should write their own procedure based on their equipment and their
quality program. Each procedure should be accompanied by an uncertainty
analysis also based on the laboratory's equipment and environment.
Figure 4-1 shows a flow chart of the calibration procedure.

Fundamentals

1. Only personnel that are qualified should calibrate the product.
2. Input panel calibration or temperature source calibration may be optional and
depends on UUT configuration and customer requirement.

Terminology

If the optional process version (-P model) was purchased, the 9190A has an additional
input panel that requires routine calibration. The optional process version (-P model) is
also referred to as the Input panel (see –P Option Panel in Chapter 2).
Table 4-10 lists the terms that are used throughout this procedure.
Term
UUT
Unit under test (for this procedure, this is the 9190A)
Test Sleeve Insert
A cylindrical aluminum insert with drilled probe holes that fits into the UUT.
Readout
Electronic device that can measure the probe resistance and convert it to
temperature.
PRT
Platinum Resistance Thermometer (also referred to as a temperature probe).
Primary PRT or
The temperature probe that is placed in the insert front hole at 0 mm. (Probe is
Primary Reference
bottomed out in the front-insert hole.) See Figure 4-2.
Secondary PRT or
The temperature probe that is placed in the insert front hole at 30 mm. The
Secondary Ref
Reference Probe bottoms out at 30 mm in the rear insert hole. Figure 4-2.
As Found Data
Data taken on the UUT before it is calibrated.
As Left Data
Data taken on the UUT after it is calibrated.
SETPOINT
Target temperature value the UUT controls to.
Axial Uniformity
The difference in temperature at one height in the UUT temperature block to a
Error or Uniformity
different height in the temperature block along the z axis. The Axial Uniformity is
Error
referred to as the Vertical Gradient.
Note
Table 4-10. Calibration Terminology
Definition
Calibration

9190A Input Panel Calibration

4
4-15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents