Fluke 9190A Operator's Manual page 78

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A
Operators Manual
SOUR:LCON:INT[?] [n]
This command is unconditionally protected, that requires a password to set it.
Read or set the main control loop integral time in seconds. The main zone integral is the
integration time in seconds that the instrument's PID controller uses for main zone control.
Range = {10.0-999.9}
Read Example:
Response:
20.0
Set Example:
SOUR:LCON:PBAN[?] [n]
This command is unconditionally protected, that requires a password to set it.
Read or set the main control loop proportional band in °C.
Range = {1.0-99.9}
Read Example:
Response:
1.5
Set Example:
SOUR:LIST:SPO<i>[?] [n]
Read or set a main temperature preset SETPOINTs.
Read example:
Response:
Set Example:
SOUR:PROT:HCUT?
Read the hard cutout temperature set-point in °C or °F. Returns the current value of the hard
cutout SETPOINT
Read Example:
Response:
140
SOUR:PROT:CLEA
Reset the cutout to enable the system. If the Calibrator exceeds the temperature set in the soft
cutout menu or if it exceeds the maximum operating temperature, a cutout condition occurs. If
this happens, the Calibrator enters cutout mode and will not actively heat or cool until the user
issues this command to clear the cutout or resets the Calibrator. See Chapter 2.
Example:
SOUR:PROT:CLEA
Response: This command has no response.
3-22
SOUR:LCON:INT?
SOUR:LCON:INT 10
SOUR:LCON:PBAN?
SOUR:LCON:PBAN 7
SOUR:LIST:SPO6?
25.00
SOUR:LIST:SPO6 100.00
SOUR:PROT:HCUT?
Note
Note