SYST:COMM:SER:LIN[?] [n]
Set serial interface linefeed enable. "n" is a value 1 or 0.
[0] = LF OFF, [1] = LF ON; Default: 1 (OFF).
Read Example:
Response:
0
Set Example:
SYST:CONF:MOD?
Read the presence of the –P module.
[0] if no –P sensor module, [1] if –P sensor card is installed
Example:
SYST:CONF:MOD?
Response:
1
SYST:DATE[?] [<year>,<month>,<day>] (-P model only)
This command is conditionally protected, that requires a password to set it.
Read or set the System Date Setting with numbers separated by commas (yyyy,mm,dd).
Read Example:
Response:
2007,05,24
Set Example:
SYST:DEC:FORM[?] [n]
Read or set the decimal format. "n" is period [0], comma [1]. Default: 0 (Period)
Read Example:
Response:
Set Example:
SYST:ERR?
Read the most resent error from the error queue. This command response reports the errors in the
error queue.
Example:
SYST:ERR?
Response:
SYST:KLOC[?] [n]
This command is unconditionally protected, that requires a password to set it.
Locks or unlocks the system keypad that restricts control to only through the serial interface
(RS-232 port) or the keypad.
[0] = unlock, and [1] = lock. Default: 0 (Unlock).
Read Example:
Response:
Set Example:
3-26
SYST:COMM:SER:LIN?
SYST:COMM:SER:LIN 1
SYST:DATE?
SYST:DATE 2007,05,24
SYST:DEC:FORM?
0
SYST:DEC:FORM 1
command protected
SYST:KLOCK?
1
SYST:KLOC 1