SOUR:PROT:SCUT:LEV[?] [n]
This command is conditionally protected, that requires a password to set it.
Read or set the soft cutout SETPOINT. The soft cutout should be set to protect the temperature
limits of the instruments under test. "n" is an integer.
Range = {-125 to 160}
Read Example:
SOUR:PROT:SCUT:LEV?
Response:
125
Set Example:
SOUR:PROT:SCUT:LEV 105
SOUR:PROT:TRIP?
Read the temperature cutout tripped state. A value of 0 is returned if the cutout set point has not
been reached. Otherwise a value of 1 is returned and the cutout set point has been reached.
Range = {0, 1}; 0 = No Cutout; 1 = Cutout.
Example:
SOUR:PROT:TRIP?
Response:
0
SOUR:RATE[?] [n]
Read or set the control temperature rate of change (Scan Rate), °C or °F per minute. The response
to this command starts out high initially and decreases as the SETPOINT is reached.
Range = {Min: 0.10, Max: 500.00} Default: 100.00.
Read Example:
SOUR:RATE?
Response:
0.531
Set Example:
SOUR:RATE 1.26
SOUR:SENS:CAL:GRAD<x>[?] [n]
This command is unconditionally protected, that requires a password to set it.
Read or set the axial gradient control parameter, where "x" is a numeric value that shows the
parameter. [1] = GRAD1 = GRAD 1, [2] = GRAD2 = GRAD 2, [3] = GRAD3. "n" is a real
number ranging from -1.0 to 1.0 entered as a ratio of the main heater power. The top zone heater
reacts as a ratio of the mains heater power to control the axial gradient.
Read Example:
SOUR:SENS:CAL:GRAD2?
Response:
0.05
Set Example:
SOUR:SENS:CAL:GRAD2 0.08
SOUR:SENS:CAL:PAR<x>[?] [n]
This command is unconditionally protected, that requires a password to set it.
Read or set a control temperature calibration parameter. "x" is a numeric value that shows the
parameter. [1] = PAR1=Temp 1,PAR2=Temp 2, PAR3=Temp 3. "n" is the entered value of the
parameter. Range = ±50.00; Defaults: 0.000. The Chapter 4 for more information.
Read Example:
SOUR:SENS:CAL:PAR1?
Response:
0.0
Set Example:
SOUR:SENS:CAL:PAR2 0.02
Note
Note
Note
Remote Operation
Remote Operation Commands
3
3-23