Control Setup - Fluke 9190A Operator's Manual

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A
Operators Manual

Control Setup

Table 2-19 shows and describes the menu selections on the Control Setup menu.
MENU
Item
TEMP PB sets the main zone proportional band and the gain in °C that the proportional-integral-
derivative (PID) controller uses for main zone control.
TEMP INT sets the main zone integral, which is the integration time in seconds that the PID
controller uses for main zone control.
TEMP DER sets the main zone derivative, which is the derivative time in seconds that the PID
controller uses for main zone control.
GRAD PB sets the gradient proportional band and the gain in °C at the proportional-integral-
derivative (PID) controller uses for gradient zone control.
GRAD INT sets the gradient integral, which is the integration time in seconds that the PID
controller uses for gradient zone control.
REF PB sets the main zone proportional band like item  when an external reference probe is
used for control.
REF INT sets the main zone integral like item  when an external reference probe is used
control.
REF DER sets the main zone derivative like item  when an external reference probe is used
for control.
2-20
Table 2-19. Control Setup Menu
S Y S T E M
M E N U
C A L I B
66.03°C
C O N T R L
TEMP DER:
Function
C O N T R O L S E T U P
TEMP PB:
10.0
TEMP INT:
100.0
8.0
GRAD PB:
20.0
GRAD INT:
150.0
REF PB:
10.0
REF INT:
100.0
REF DER:
8.0
1
2
3
4
5
6
7
8
gzs020.eps