Fluke 9190A Operator's Manual page 75

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

PROG:IDEN[?] [n]
Read or set the program identifier.
Character range = {0 – 9, A – Z, '-'}, up to 12 characters, minimum 1 character: Default: "0".
Read Example:
PROG:IDEN?
Set Example:
TEST-1
PROG:MEM:CLE (-P Model only)
This command is unconditionally protected, that requires a password to set it.
Erase all test reports stored in NVMemory.
Example:
PROG:MEM:CLE
PROG:MEM:COUN? (-P Model only)
Read the test report count. Provide the count for the number of test reports currently stored in
memory.
Example:
PROG:MEM:COUN? 6
PROG:MEM:PRIN [n] [ALL] (-P Model only)
Prints one or all test reports. "n" indicates the test report to be printed and 1 is the earliest test.
Example:
PROG:MEM:PRINT 1
PROG:SEQ:CAT?
This command provides a list of the parameters for the Ramp and Soak test.
Example:
PROG:SEQ:CAT?
Response:
"SPOn","DWELL","DIR","POIN","CYCL","PTOL"
PROG:SEQ:PAR? par[,<n>]
Read or set a program parameter for Ramp & Soak tests.
Range = {SPOn, DWELL, DIR, POIN, CYCL, PTOL}.
Set Example:
PROG:SEQ:PAR cycle,8
Parameter
SPO[n]*
DWEL
POIN
CYCL
PTOL
DIR
*Read Only, must be <= # of setpoints (POIN) Read Example: PROG:SEQ:PAR? dwell Response: 25
Note
Table 3-5. PROG:SEQ:PAR Parameters
Minimum
1
1
1
1
0.01
0 (up)
Remote Operation Commands
Maximum
8
100
8
999
99.9
1 (U/D)
Remote Operation
Default
1
15
8
1
1.00
0
3-19
3