PROG:STAT[?] [0|1]
Read or set the execution state for the selected Program (Off=0, Run=1, Default = 0). If the
selected program is not running then a value of 0 is returned otherwise a 1 is returned.
Read Example:
Response:
Set Example:
PROG:TYPE[?] [ <prog>]
Read or select a program to run where "prog" is a name, SEQ, ASW, or MSQ.
Default = SEQ.
Read Example:
Response:
SEQ
Set Example:
READ?, MEAS? or FETC? (-P model only)
Read the Reference sensor temperature, °C or °F. If the external reference probe is enabled, the
reference temperature is returned otherwise, 0.0 is returned.
Example:
READ?
Response:
264.262
SENS1:CAL:PAR<n>[?][cal] (-P model only)
This command is unconditionally protected, that requires a password to set it.
Read or set a reference input calibration parameter. "n" is a value of 1 or 2 corresponding to the
calibration parameters REF1C0 and REF1C100 respectively. "cal" is a real number used as the
calibration offset for the respective parameter.
REF1C0 Range = {-1.0 to 1.0} REF1C100
Range = {-2.0 to 2.0} Defaults (all): 0.0000
Read Example:
Response:
0.2
Set Example:
SENS1:DATA? (-P model only)
Read the reference input resistance. This command returns the resistance in ohms of the reference
probe.
Example:
SENS1:DATA?
Response:
199.9366
3-20
PROG:STAT?
0
PROG:STAT 1
PROG:TYPE?
PROG:TYPE ASW
SENS:CAL:PAR1?
SENS1:CAL:PAR2 0.092