8. Calculate the error with the formula below where actual is the calibrated resistance
value of the standard resistor. Record this result.
Error = Average – Actual
9. Compare the error with the spec to determine the status. Record the result.
Resistance Standard
Resistance Standard
UUT PRT 3-Wire Test
The UUT PRT 3-Wire Test is used to collect As Found and As Left data.
All Reference Accuracy Tests are to be performed in Ohms.
Perform the UUT PRT 3-Wire Test as follows:
1. Connect the 100 Ω resistor in the 3-wire UUT configuration (see Figure 4-3).
2. Set the instrument to take UUT measurements from the PRT input.
3. Set the instrument to 3-wire mode for the UUT PRT input.