Fluke 9190A Operator's Manual page 73

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

CALC1:CONV:SNUM[?] <ser_num> (-P model only)
This command is unconditionally protected, that requires a password to set it.
Read or enter a reference probe serial number. "ser_num" character range equals {0-9, A-Z, '-'},
up to 10 characters in length with a minimum of 1 character length. Default: "0"
When a blank space is entered, any characters after the blank are dropped. For example, change
S/N 1234-5678 to S/N TEST1. Enter TEST1<Blank Space>678. The serial number will drop the
last three characters and enter the S/N TEST1.
Read Example:
CALC1:CONV:SNUM?
Response:
1234
Set Example:
CALC1:CONV:SNUM 1560-D
CALC1:CONV:TEST[?] <n>]
Test the conversion algorithm from resistance (ohms) to temperature (°C or °F). "n" is the value
to be converted (ohms). Entry for "n" is required for converted output.
Range = {0-500};
Default = 100.
Read Example:
CALC1:CONV:TEST? 100.000
Response:
0.0100
CAL1:DATA? (-P model only)
Reads the reference sensor Temperature. The returned value will be in degrees °C (°F) if a
temperature value is returned. The value can be a resistance if the conversion selection is
RESISTANCE.
Example:
CALC1:DAT?
Response:
325
CALC2:CONV:CAT? <func> (-P model only)
Read the list of UUT probe characterization names. "func" is the UUT device. "func" is not
required and defaults to the selected device based on the setting of <SENS2:FUNC>.
Range = {RTD, TC, MA},
Default = RTD (if no previous selection), else current selection.
Example 1:
CALC2:CONV:CAT? TC
Response 1:
"C","E","J","K","L","M","N","R","S","T","U","mV"
If SENS2:FUNC is set to RTD and the command is sent without the function:
Example 2:
CAL2:CONV:CAT?
Response 2:
"PT_A385","PT_A392","PT_JIS","NI_120","RES"
CALC2:CONV:NAME[?] n] (-P model only)
Read or set the UUT device characterization scheme. Must be consistent with the selected
Function and device. For RTD: "n" is the RTD type as follows: PT A385; PT_A392; PT_JIS;
NI_120; or RES. For TC; "n" is the thermocouple type as follows: C; E; J; K; L; M; N; R; S; T;
U; or mV. Default: RTD, PT_A385. Command responds with the selected UUT probe
characterization method; RES, I90, IEC, RES, mV, or mA.
Read Example:
CALC2:CONV:NAME?
Response:
RTD
Set Example:
CALC2:CONV:NAME NI_120
Note
Remote Operation
Remote Operation Commands
3
3-17