Terminology - Fluke 9190A Operator's Manual

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

Terminology

Table 4-1 lists the terms that are used throughout this procedure.
Term
UUT
Unit under test (for this procedure, this is the 9190A)
Test Sleeve Insert
A cylindrical aluminum insert with drilled probe holes that fits into the UUT.
Readout
Electronic device that can measure the probe resistance and convert it to
temperature.
PRT
Platinum Resistance Thermometer (also referred to as a temperature probe).
Primary PRT or
The temperature probe that is placed in the insert front hole at 0 mm. (Probe is
Primary Reference
bottomed out in the front-insert hole.) See Figure 4-2.
Secondary PRT or
The temperature probe that is placed in the insert rear hole at 30 mm up from the
Secondary Ref
bottom of the Insert. The Reference Probe bottoms out at 30 mm in the rear insert
hole. Figure 4-2.
As Found Data
Data taken on the UUT before it is adjusted.
As Left Data
Data taken on the UUT after it is adjusted.
SETPOINT
Target temperature value the UUT controls to.
Axial Uniformity
The difference in temperature at one height in the UUT temperature block to a
Error or Uniformity
different height in the temperature block along the vertical axis.
Error
Table 4-1. Calibration Terminology
Definition
Calibration
Temperature Source Calibration
4
4-5