Set Temperature; Set The Temperature Setpoint Manually; Set The Temperature With A Preset Setpoint - Fluke 9190A Operator's Manual

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A
Operators Manual

Set Temperature

A set temperature value is referred to as a "SETPOINT". The Calibrator uses the
SETPOINT value to know what temperature to heat or cool to. Use the procedures in the
subsequent sections to manually set a SETPOINT, load a SETPOINT preset, or change a
SETPOINT preset.

Set the Temperature SETPOINT Manually

To adjust the Calibrator temperature SETPOINT manually (see Figure 2-10):
1. Push .
2. Push .
3. Push  and  to highlight a digit then push  to increase the digit or push  to
4. Push  twice to save the SETPOINT temperature and start the Calibrator.

Set the Temperature with a Preset SETPOINT

The Calibrator stores eight user-configured preset SETPOINT temperatures to help
quickly set a temperature. To change a SETPOINT preset see "Change a SETPOINT
Preset" on page 2-37.
To load and use a preset SETPOINT (see Figure 2-11):
1. Push  on the Main screen.
2. Push .
3. Push  or  to move between presets (1 through 8).
4. Push  to select a preset.
2-36
decrease the digit.
S E L E C T
P R E S E T
Figure 2-10. Set Temperature SETPOINT
E D I T
P R E S E T
Figure 2-11. Preset Selection
66.03°C
S E T T E M P E R A T U R E S E T P O I N T
SETPOINT: +080.00
°C
S A V E /
D I S A B L E
66.03°C
S E L E C T P R E S E T
PRESET:
(SETPOINT: 80.00 °C)
gzs035.eps
gzs034.eps