Fluke 9190A Operator's Manual page 80

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A
Operators Manual
SOUR:SENS:CAL:TEMP<x>?
Read the required calibration temperature (°C) that corresponds to a calibration parameter. "x" is
a numeric value indicating the parameter [1] = TEMP1, [2] = TEMP2, AND [3] = TEMP3.
Range = {1-3}; Default = 1.
Example:
SOUR:SENS:CAL:TEMP1?
Response:
40
SOUR:SENS:DATA? [TEMP]
Read the control temperature, °C or °F. The current control temperature is returned if the above
or if TEMP is appended to the end of the example.
Example:
SOUR:SENS:DATA?
Response:
30.285°C
SOUR:SENS:DATA? [RES]
Read the control sensor resistance. When RES is appended to the end of the example above, the
internal sensor resistance is returned.
Example:
SOUR:SENS:DATA? RES
Response:
SOUR:SENS:ROUT {0 | 1}
Enable/disable the external control sensor.
SOUR:SPO[?] [n]
This command reads or sets the value of the control SETPOINT in the temperature unit set.
Range = {Min -110.00, Max 140.00} Default = 25.00
Read Example:
Response:
50.000
Set Example:
SOUR:STAB:BEEP[?] [n]
Enable or disable the audible stability alert.
[0] is disable, [1] is enable beep.
Default:1 (Enable Beep).
Read Example:
Response:
1
Set Example:
SOUR:STAB:DAT?
Read the control temperature stability, °C or °F. The controller stability is returned.
Example:
SOUR:STAB:DAT?
Response:
0.306
3-24
or
(current control temp)
111.28
SOUR:SPO?
SOUR:SPO 100.00
SOUR:STAB:BEEP?
SOUR:STAB:BEEP 0
SOUR:SENS:DATA? TEMP