8 – Language Dictionary
[SOURce:]VOLTage:PROTection:STATe
This command enables or disables the automatic overvoltage protection tracking (OVP) function. It does
not disable the programmable VOLTage:PROTection level.
CAUTION:
Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are disconnected, to damage the equipment under test.
Command Syntax [SOURce:]VOLTage:PROTection:STATe <bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples VOLT:PROT:STAT 0
Query Syntax [SOURce:]VOLTage:PROTection:STATe?
Returned Parameters <NR1>0 or 1
Related Commands OUTP:PROT:CLE
[SOURce:]VOLTage:TRIGger
This command sets the pending triggered voltage level of the dc source. The pending triggered level is a
stored voltage value that is transferred to the output terminals when a trigger occurs. In order for a trigger
to occur, the trigger subsystem must be initiated (see the INITiate command in the trigger subsystem).
Command Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
Parameters see Table 8-3
Default Suffix V (volts)
*RST Value 0
Examples VOLT:TRIG 20
Query Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Returned Parameters <NR3>
Related Commands VOLT
[SOURce:]VOLTage2:TRIGger
Agilent 66319B/66319D only
This command sets the triggered voltage level of the auxiliary output. The triggered level is a stored
value that is transferred to the output when a trigger occurs. In order for a trigger to occur, the trigger
subsystem must be initiated (see the INITiate command).
Command Syntax [SOURce:]VOLTage2[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
Parameters see Table A-3
Default Suffix V (volts)
*RST Value 0
Examples VOLT2:TRIG 20
Query Syntax [SOURce:]VOLTage2[:LEVel]:TRIGgered[:AMPLitude]?
Returned Parameters <NR3>
Related Commands VOLT2
116
VOLT:PROT:STAT 1
*RST
VOLT:PROT
VOLT2:PROT
!voltage protection is OFF
!voltage protection is ON
VOLT:PROT
VOLTAGE:LEVEL:TRIGGERED 20
VOLT2:LEV:TRIG 20