Power-On Conditions
Refer to the *RST command description in chapter 8 for the power-on conditions of the status registers.
QUESTIONABLE STATUS
CONDITION
PTR/NTR
0
1
1
OV
1
2
2
OC
2
N.U.
3
FP
8
8
4
OT
16
16
5
OS
32
32
6-7
N.U.
8
256
256
UNR2
9
512
512
RI
10
UNR
1024
1024
11
N.U.
12
OC2
4096
4096
13
N.U.
14
OVLD
16384
16384
15
N.U.
STANDARD EVENT STATUS
EVENT
ENABLE
0
OPC
1
1
1
N.U.
2
QYE
4
4
3
DDE
8
8
4
EXE
16
16
5
CME
32
32
6
N.U.
7
128
128
PON
OPERATION STATUS
CONDITION
PTR/NTR
0
CAL
1
1
1-4
N.U.
5
WTG
32
32
6,7
N.U.
8
CV
256
256
9
512
512
CV2
10
1024
1024
CC+
11
2048
2048
CC-
12
4096
4096
CC2
13-15
N.U.
EVENT
ENABLE
1
1
2
2
8
8
16
16
32
32
256
256
LOGICAL
OR
512
512
1024
1024
4096
4096
16384
16384
OUTPUT QUEUE
DATA
DATA
DATA
LOGICAL
OR
EVENT
ENABLE
1
1
32
32
256
256
LOGICAL
OR
512
512
1024
1024
2048
2048
4096
4096
Figure 7-8. DC Source Status Model
Programming the DC Source - 7
OFF
N.U.
QUEUE
QUES
NOT
EMPTY
MAV
ESB
MSS
OPER
OUTPut:DFI
:SOURce
SERVICE
REQUEST
ENABLE
STATUS BYTE
0-2
3
8
8
4
16
16
5
32
32
6
64
7
128
128
RQS
SERVICE
REQUEST
GENERATION
FLT
LOGICAL
OR
FIG4-6.GAL