Triggering Output Changes - Agilent Technologies 66319B User Manual

Mobile communications dc source
Hide thumbs Also See for 66319B:
Table of Contents

Advertisement

Triggering Output Changes

The dc source has two independent trigger systems. One is used for synchronizing output changes, and
the other is used for synchronizing measurements. This section describes the output trigger system. The
measurement trigger system is described under "Triggering Measurements".
SCPI Triggering Nomenclature
In SCPI terms, trigger systems are called sequences. When more than one trigger system exists, they are
differentiated by naming them SEQuence1 and SEQuence2. SEQuence1 is the transient trigger system
and SEQuence2 is the measurement trigger system. The dc source uses aliases with more descriptive
names for these sequences. These aliases can be used instead of the sequence forms.
Sequence Form
Alias
SEQuence1
TRANsient
SEQuence2
ACQuire
Output Trigger Model
Figure 7-1 is a model of the output trigger system. The rectangular boxes represent states. Arrows show
the transitions between states. These are labeled with the input or event that causes the transition to
occur.
INITiate:CONTinuous OFF
INITiate:CONTinuous ON
Setting the Voltage, Current, or Resistance Transient Levels
To program output trigger levels, you must first specify a voltage or current trigger level that the output
will go to once a trigger signal is received. Use the following commands to set the output trigger level:
VOLT:TRIG
<n>
VOLT2:TRIG <n>
CURR:TRIG
<n>
CURR2:TRIG <n>
RES:TRIG
<n>
NOTE:
Until they are programmed, trigger levels will be the same as the corresponding voltage
or current levels. For example, if a dc source is powered up and the voltage is
programmed to 6, the trigger level is also set to 6. Once you program a trigger level, it
will stay at that value until the output is changed by a transient trigger or reprogrammed.
Figure 7-1. Model of Output Trigger System
for models that have a second output
for models that have a second output
only applies to output 1 (the main output)
Programming the DC Source - 7
IDLE STATE
INITiate[:IMMediate]
INITIATED STATE
TRIGGER RECEIVED
OUTPUT
LEVEL
CHANGE
ABORt
*RST
*RCL
71

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

66319d66321b66321d

Table of Contents