Agilent Technologies 66319B User Manual page 135

Mobile communications dc source
Hide thumbs Also See for 66319B:
Table of Contents

Advertisement

A serial poll also returns the value of the Status Byte register, except that bit 6 returns Request for
Service (RQS) instead of Master Status Summary (MSS). A serial poll clears RQS, but not MSS. When
MSS is set, it indicates that the dc source has one or more reasons for requesting service.
Bit Position
7
Bit Name
OPER
Bit Weight
128
ESB = Event status byte summary
MAV = Message available
MSS = Master status summary
Query Syntax *STB?
Returned Parameters <NR1>(Register binary value)
*TRG
This common command generates a trigger when the trigger subsystem has BUS selected as its source.
The command has the same affect as the Group Execute Trigger (<GET>) command.
Command Syntax *TRG
Parameters None
Related Commands ABOR INIT TRIG[:IMM]
*TST?
This query causes the dc source to do a self-test and report any errors. 0 indicates that the dc source
passed self-test. 1 indicates that one or more tests failed. Selftest errors are written to the error queue (see
Appendix C).
Query Syntax TST?
Returned Parameters <NR1>
*WAI
This command instructs the dc source not to process any further commands until all pending operations
are completed. "Pending operations" are as defined under the *OPC command. *WAI can be aborted
only by sending the dc source an GPIB DCL (Device Clear) command.
Command Syntax *WAI
Parameters None
Related Commands *OPC* OPC?
Table 8-8. Bit Configuration of Status Byte Register
6
5
MSS
ESB
MAV
(RQS)
64
32
OPER = Operation status summary
QUES = Questionable status summary
RQS = Request for service
4
3
2
QUES
0
16
8
4
<GET>
Language Dictionary - 8
1
0
0
0
2
1
135

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

66319d66321b66321d

Table of Contents