Agilent Technologies 66319B User Manual page 40

Mobile communications dc source
Hide thumbs Also See for 66319B:
Table of Contents

Advertisement

4 - Turn-On Checkout
Procedure
7.
Plug the output connector
back into the unit.
8.
Press Shift, Prot Clear
9.
Press Voltage
10.
Press Enter Number,
<15>, Enter
11.
Press Output On/Off
12.
Connect a jumper wire
across the + and - output
terminals.
Press Output On/Off.
13.
14.
Press Current,
Enter Number,
<1>, Enter.
Press Shift, OCP
15.
16.
Press Shift, OCP
Press Shift, Prot Clear
17.
18.
Turn the unit off and remove
the shorting wire from the
output terminals.
Only perform steps 19 to 29 if you are verifying an Agilent 66319B or 66319D unit.
Procedure
19.
Turn the unit on. Wait for
selftest to complete and
press Shift, Channel.
20.
Press Voltage,
Enter Number,
<12>, Enter.
21.
Press Output On/Off
22.
Press Output On/Off
40
Display
Explanation
Restores the output sense connections. The Prot
annunciator is still on.
NO FAULT
Clears the protection condition. Prot is off; CV is on.
VOLT 0.000
Display shows the output voltage setting of the unit.
VOLT <15>
Programs the main output to 15 volts. After the value
15.003V 0.0001A
is entered, the display returns to Meter mode.
Because the output is enabled, the meter will indicate
the actual output voltage.
0.000V
0.0000A
Turn the output off.
Shorts the output of the unit.
0.004V 3.0712A
The CC annunciator is on, indicating that the unit is
in constant current mode. The unit is sourcing output
current at the maximum rating, which is the default
output current limit setting.
CURR <1>
Programs the output current to 1 ampere.
0.001V 0.0003A
You enabled the overcurrent protection circuit. The
circuit then tripped because the unit was operating in
constant current mode. The CC annunciator turns off,
and the OCP and Prot annunciators turn on.
0.001V 0.0003A
You have disabled the overcurrent protection circuit.
The OCP annunciator turns off.
0.004V
0.998A
Restores the output. The Prot annunciator turns off.
The CC annunciator turns on.
The next time the unit turns on it will be restored to
the *RST or factory default state.
Display
Explanation
2
0.025V 0.0002A
Shift Channel toggles between channel 1 and channel
2. The left-most digit of the display identifies the
output channel that is presently being controlled. It will
indicate a "1" for channel 1, or "2" for channel 2.
2
VOLT <12>
Programs the output 2 voltage to 12 volts.
2
12.005V 0.0002A
Turns the main output and output 2 on. The Dis
annunciator is off, but the CV annunciator is on.
2
0.000V
0.0000A
Turn all outputs off.