8 – Language Dictionary
SENSe:LEAD:STATus?
This query returns the status of the open sense detection circuit. The query must be performed with the
output disabled. Any external source such as an external capacitor must be discharged. The following
status code is returned:
Value
Description
0
Sense lead connections are normal
Positive sense lead is open
1
2
Negative sense lead is open
3
Bothe sense leads are open
Query Syntax SENSe:LEAD:STATus?
Returned Parameters <NR3>
Examples SENS:LEAD:STAT?
SENSe:FUNCtion
This command configures the sensing function for triggered measurements. The dc source has up to three
measurement sensors as described below. The query returns the function setting.
CURRent
Senses the current measurement at the main output (output 1)
Senses the voltage measurement at the DVM inputs (Agilent 66321D/66319D only)
DVM
VOLTage
Senses the voltage measurement at the main output (output 1)
Command Syntax SENSe:FUNCtion <function>
Parameters "VOLTage" | "CURRent" | "DVM"
Examples SENS:FUNC "VOLT"
Query Syntax SENSe:FUNCtion?
Returned Parameters <SRD>
SENSe:PROTection:STATe
This command enables or disables the open sense lead detection. The default setting for this function is
disabled. To permanently enable open sense lead detection, program this command on, save this state in
location 0 using *SAV 0, and set the output power-on state to recall state 0 (OUTP:PON:STAT RCL 0).
Command Syntax SENSe:PROTection:STATe <bool>
Parameters 0 | OFF | 1 | ON
*RST Value 0
Examples SENS:PROT:STAT ON
Query Syntax SENSe:PROTect:STATe?
Returned Parameters <NR3>
SENSe:SWEep:OFFSet:POINts
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger. Positive values represent the delay after the trigger
occurs but before the samples are acquired.
Command Syntax SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters -4095 through 2,000,000,000
*RST Value 0
Examples SENS:SWE:OFFS:POIN -2047
Query Syntax SENSe:SWEep:OFFSet:POINts?
Returned Parameters <NR3>
Related Commands SENS:SWE:TINT SENS:SWE:POIN
106
Value
Description
4
Sense lead status is unknown, possibly
caused by an external voltage
5
The output is not in the disabled state
MEAS:ARR