Apg Voltage Calibration - Chroma 62000H Series Operating & Programming Manual

Programmable dc power supply
Hide thumbs Also See for 62000H Series:
Table of Contents

Advertisement

S E T T I N G
A C T U A L
12. The current calibration is done once the above actions are completed and the DC Power
Supply output will be turned off. To save the calibration data, press "
a confirmation page as Figure 3-124 shows. Press "
SAVE=YES and press "
to return to the Calibration screen.
13. Press "
" to return to the MAIN PAGE.
EXIT
WARNING
Improper shunt range selection may cause damage to the current shunt.
The calibration point may be different for other models (non
62150H-600S), please operate it following the instructions displayed.
3.3.7.4

APG Voltage Calibration

3.3.7.4.1
Hardware Requirements
Table 3-9 lists the hardware requirements for APG voltage calibration.
Device
DVM
DC Power Supply Any DC Power Supply or DC signal source
[ C U R R E N T
O U T P U T
C U R R E N T = 1 9 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0 _
Figure 3-123
" to save it. If there is no need to save it, press "
ENTER
S A V E = N O
Figure 3-124
Table 3-9
Suggest Model or Capacity
HP 34401A or equivalent DVM
that can output 10Vdc and drive 100mA.
S E T T I N G
C A L I B R A T I O N ]
1
" or "Rotary" (
[ S A V E A R G U M E N T ]
Manual Operation
A
[ 4 ]
A
" will prompt
SAVE
) to set
EXIT
3-69
"