Chroma 62000H Series Operating & Programming Manual page 122

Programmable dc power supply
Hide thumbs Also See for 62000H Series:
Table of Contents

Advertisement

Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
A P P L Y
S E T
L O A D I N G
A C T U A L O U T P U T
Next, press "
9.
as shown in Figure 3-146, set the Electronic Load to load 36.00A current. Enter the
current read by the ULTRASTAB SATURN (DVM) to position [4] and press "
confirm it. Using the calibration of 12.00A and 36.00A, the system will calculate the
calibration factor of high current range.
A P P L Y
S E T
L O A D I N G
A C T U A L O U T P U T
10. The current calibration is done when the above actions are completed. To save the
calibrated value, press "
Figure 3-147. Press "
press "
" to save it. If not saving it, press "
ENTER
11. Press "
" to return to the MAIN PAGE.
EXIT
3-80
[C U R R E N T
L O A D I N G A N D P R E S S[ E N T E R ]
C U R R E N T
C U R R E N T= 0 . 0 0 0 0 _ A
Figure 3-145
" to do 36.00A calibration. When the cursor stops at position [4]
[C U R R E N T
L O A D I N G A N D P R E S S[ E N T E R ]
C U R R E N T
C U R R E N T= 0 . 0 0 0 0 _ A
Figure 3-146
" and a confirmation screen will appear as shown in
SAVE
1
" or turn the "Rotary" (
S A V E = N O
Figure 3-147
M E A S
C A L I B R A T I O N ]
= 1 2 . 0 0 0
A
M E A S
C A L I B R A T I O N ]
= 3 6 . 0 0 0
A
) knob to set SAVE=YES and
" to return to Calibration page.
EXIT
[ S A V E A R G U M E N T ]
[ 3 ]
" to
ENTER
[ 4 ]