Dismantling the swinging axle:
·
Heat the screw (3) using a hot air gun to soften the thread locking adhesive.
·
Remove the screw (3) with a 10 mm Allen key.
·
Pull the swinging axle (1) off the chassis and remove the guide bushes (2 and 4).
Refitting the swinging axle:
·
Refit the parts in reverse order.
·
Tighten the screw (3) to 70 Nm +/- 10 Nm.