PROBLEM
Mobility
device
judders in
drive mode
Batteries
not being
charged
Mobility
device runs
too slowly
20
OTHER
POSSIBLE
SYMPTOMS
CAUSE
None
Batteries
defective
(unstable
voltage)
Drive motor(s)
defective
None
Batteries
defective
LEDs blinking
Charging unit
on charging
defective
unit
None
Remote
defective
Batteries
defective
SOLUTION
DOCUMENTATION
Replace batteries
See chapter 7.4.
Replace motor(s)
See chapter 7.11.1.
Replace batteries
See chapter 7.4.
Replace charging
See charging unit
unit
user manual
Replace remote
See chapter 7.2.2.
Replace batteries
See chapter 7.4.
®