Table 6-2 Performance Verification Tests - Agilent Technologies U1253B User's And Service Manual

Agilent u1253b true rms oled multimeter
Hide thumbs Also See for U1253B:
Table of Contents

Advertisement

Table 6-2 Performance verification tests

Step
Test function
1
Turn the rotary switch to the
2
Press
to switch to frequency mode
3
Press
to switch to duty cycle mode
4
Turn the rotary switch to the
Press
to select DC V measurement
U1253B User's and Service Guide
Range
[1]
position
5 V
50 V
500 V
1000 V
9.9999 kHz
0.01% to
99.99%
position
5 V
50 V
500 V
1000 V
Performance Tests and Calibration
Reference
signals/values
Error limits
5520A output
5 V, 1 kHz
± 22.5 mV
5 V, 10 kHz
± 79.0 mV
4.5 V, 20 kHz
± 169.5 mV
4.5 V, 30 kHz
± 169.5 mV
4.5 V, 100 kHz
± 169.5 mV
50 V, 1 kHz
± 225.0 mV
50 V, 10 kHz
± 790.0 mV
45 V, 20 kHz
± 1.695 V
45 V, 30 kHz
± 1.695 V
45 V, 100 kHz
± 1.695 V
500 V, 1 kHz
± 2.25 V
1000 V, 1 kHz
± 8.0 V
0.48 V, 1 kHz
± 500 mHz
5.0 Vpp @ 50%,
± 0.315%
square wave, 50 Hz
5 V
± 1.75 mV
50 V
± 17.5 mV
500 V
± 200 mV
1000 V
± 800 mV