Duty cycle
U1253B User's and Service Guide
[1]
and pulse width
Table 7-11 Accuracy for duty cycle
Mode
DC coupling
Table 7-12 Accuracy for pulse width
Range
500 ms
2000 ms
[1]
The accuracy for duty cycle and pulse width is based on a 5 V square wave input
into the DC 5 V range. For AC coupling, the duty cycle range can be measured
within 5% to 95% for signal frequency > 20 Hz.
Positive or negative pulse width must be greater than 10 μs and the range of
[2]
duty cycle should be considered. The range of pulse width is determined by the
frequency of the signal.
[2]
Range
0.01% to 99.99%
Resolution
0.01 ms
0.1 ms
Specifications
Accuracy of full scale
0.3% per kHz + 0.3%
Accuracy
0.2%+3
0.2%+3
185