Table 6-4 List Of Calibration Items - Agilent Technologies U1253B User's And Service Manual

Agilent u1253b true rms oled multimeter
Hide thumbs Also See for U1253B:
Table of Contents

Advertisement

Table 6-4 List of calibration items

Function
AC V
(with rotary switch at
[2]
and
)
DC V
DC mV
U1253B User's and Service Guide
Range
Calibration item
5 V
0.3000 V (1 kHz)
3.0000 V (1 kHz)
3.0000 V (10 kHz)
50 V
3.000 V (1 kHz)
30.000 V (1 kHz)
30.000 V (10 kHz)
500 V
30.00 V (1 kHz)
300.00 V (1 kHz)
300.00 V (10 kHz)
1000 V
30.0 V (1 kHz)
300.0 V (1 kHz)
300.0 V (10 kHz)
(done for this function; change
rotary switch position or press
to select the next function
that requires calibration)
Short
SHORT
5 V
3.0000 V
50 V
30.000 V
500 V
300.00 V
1000 V
1000.0 V
(done)
Short
SHORT
50 mV
30.000 mV
500 mV
300.00 mV
1000 mV
1000.0 mV
(done)
Performance Tests and Calibration
[1]
Reference input
0.3 V, 1 kHz
3 V, 1 kHz
3 V, 10 kHz
3 V, 1 kHz
30 V, 1 kHz
30 V, 10 kHz
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
Dual banana shorting plug with
copper wire
3 V
30 V
300 V
1000 V
Dual banana shorting plug with
copper wire
30 mV
300 mV
1000 mV
6
167

Advertisement

Table of Contents
loading